Specifications
Test Items WBC, LY#, MO#, GR#, LY%, MO%, GR%, HGB, RBC,
HCT, MCV, MCH, MCHC, RDWCV, RDWSD, PLT, PCT, MPV, PDW, WBC
Histogram, RBC Histogram, PLT Histogram
Background Count WBC: ≤0.3×109/L RBC: ≤0.03
×1012/L
HGB: ≤1g/L PLT: ≤10×109/L
Linearity Test Item Test Range Allow Difference of
Linearity
WBC 0~6.0×109/L ±0.3×109/L
6.0~99.9×109/L ±5%
RBC 0~0.99×1012/L ±0.05×1012/ L
1.0~9.99×1012/L ±5%
HGB 0~99 g/L ±2g/L
100~300 g/L ±2%
PLT 0~99×109/L ±10×109/L
100~999×109/L ±10%
Repeatability Test Item CV(%) Teat Range
WBC ≤2 7.015.0×109/L
RBC ≤0.5 3.56.0×1012/L
HGB ≤1.5 110180g/L
PLT ≤4 150500×109/L
MCV ≤0.5 80110fL
Sample Volume Twig blood: 20μl
Dilution Ratio WBC/HGB: 1:500 RBC/PLT: 1:50000
Interfaces RS232 joint, USB, PS2, Parallel port
Test Speed 60 samples/hours
Continuous Work Time 8h